Major II Boyle’s Machine
The Major II Boyle’s Machine is a semi-automatic anesthesia workstation that combines traditional design elements with modern safety features. It is engineered to deliver a controlled mixture of medical gases—typically oxygen (O₂) and nitrous oxide (N₂O)—to patients during surgical procedures. The machine operates on a continuous-flow principle, ensuring a steady supply of anesthetic gases.
Key Features :
- Dual Gas Delivery: Rotameters for O₂ and N₂O allow precise control over gas flow rates.
- Goldman-Type Halothane Vaporizer: Facilitates administration of halothane anesthetic agents.
- Emergency Oxygen Flush Valve: Provides rapid oxygen supply in emergencies.
- Non-Return and Pressure Relief Valve: Prevents backflow and protects the system from excessive pressure.
- Blow-Off Valve: Activates when pressure exceeds 200 cm H₂O for patient safety.
- Ventilator and CO₂ Absorber Compatibility: Allows integration with additional equipment.
- Cylinder Storage: Extended rear platform accommodates two 'B' type cylinders.
- Stainless Steel Construction: Ensures durability and ease of maintenance.
- Storage Drawer: Provides space for essential accessories and equipment.